Manual Mode; Starting A Centrifugation Run; Interrupting A Centrifugation Run; Interrupting A Deceleration Process - Sigma 2-16KL Operating Manual

Refrigerated centrifuge
Table of Contents

Advertisement

Pos : 124 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/Spinc ontrol L (Z ent6) /060- 0030-0020 M anueller Betri eb @ 27\m od_1405319254192_68.docx @ 192647 @ 3 @ 1
6.3.2

Manual mode

Pos : 125 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/Spinc ontrol L (Z ent6) /060- 0030-0020- 0010 Starten ei ner Z entrifugati on @ 27\m od_1405319255424_68.docx @ 192661 @ 4 @ 1
6.3.2.1

Starting a centrifugation run

Pos : 126 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 127 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/Spinc ontrol L (Z ent6) /060- 0030-0020- 0020 U nterbr ec hen einer Z entrifug ation @ 27\mod_1405319256724_68.doc x @ 192675 @ 4 @ 1
6.3.2.2

Interrupting a centrifugation run

Pos : 128 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 129 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/Spinc ontrol L (Z ent6) /060- 0030-0020- 0030 U nterbr ec hen eines Brems vorgangs @ 27\m od_1405319258018_68.docx @ 192689 @ 4 @ 1
6.3.2.3

Interrupting a deceleration process

Pos : 130 /010 Univ ers almodul e/
Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 131 /100 Sigma/100 BA Z entrifugen Sigma (Standardm odul e)/060 Betrieb_Spi nc ontr ol L+S/Spinc ontrol L (Z ent6) /060- 0030-0020- 0040 Auswahl, Anzeig e und Änderung von D aten @ 27\m od_1405319259317_68.docx @ 192703 @ 4 @ 1
6.3.2.4

Selection, display, and modification of data

Pos : 132 /010 Univ ers almodul e/
Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1
Version 08/2012, Rev. 1.24 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0700702
The centrifuge is ready for operation when the start key is illuminated.
Press the start key in order to start a centrifugation run.
Press the stop key in order to interrupt a centrifugation run. The
centrifugation run will be terminated prematurely.
Quick stop
Press the stop key for more than two seconds.
The centrifuge decelerates with the maximum deceleration curve.
After a quick stop, the centrifuge lid must be opened before a new
centrifugation run can be started.
A quick stop can also be triggered during a normal deceleration, e.g. in
order to speed up the deceleration.
When a quick stop is triggered, "fast" will be displayed in the lower right-
hand corner of the display.
Press the start key during a deceleration process in order to interrupt it
and to restart the centrifuge.
Turn the function knob in order to select a field. The selected field will
be preceded by the indication "set", "run", or "save".
Press the function knob. The indication ("set", "run", or "save") flashes
and the modification mode is active.
Turn the function knob in order to modify the set value of the selected
field.
Press the function knob again to confirm the input and to quit the
modification mode
The process will be interrupted if the stop key is pressed or after 60
seconds without any further input.
Refrigerated Centrifuge Sigma 2-16KL
6 Using the centrifuge
37 / 82