Transmitter Electromagnetic Immunity Specifications
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
Receiver Product Specifications
Part Number
Reading Frequency
Dimensions
TX/RX Frequencies
Bandwidth
Maximum Output Power
Modulation
Data Rate
Dexcom G5 Mobile System User Guide
Technical Information
IEC 60601 Test
Level
± 8 kV Contact
± 15 kV Air
30 A/m
Transmitter
Compliance
Level
± 8 kV Contact
± 15 kV Air
30 A/m
MT22949
Every 5 minutes
Length: 4.0 inches
Width: 2.5 inches
Thickness: 0.55 inches
2.402-2.480 GHz
1.22 MHz
2.5 mW EIRP
Gaussian Frequency-Shift Keying
1 Mbps
Electromagnetic
Environment Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.