Calibration Stability
The system must be calibrated every 12 hours. To demonstrate performance
of the system over a 12-hour calibration period, systems were evaluated to
verify that performance remains consistent over the 12-hour calibration period.
Systems were evaluated in 2-hour increments after calibration. Performance
was estimated at each 2-hour interval and stratified by glucose values by
calculating the percentage of system readings within 15 mg/dL or 15%, 20
mg/dL or 20%, 30 mg/dL or 30%, 40 mg/dL or 40% and greater than 40 mg/dL
or 40% of the YSI values in Table 7-A and 7-B.
Table 8-A. Sensor Stability Relative to YSI (Accuracy Over Time
Day
2
of
Study
Wear
Original
Day
1
Software 505
Original
Day
4
Software 505
Original
Day
7
Software 505
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number
Mean
of
Absolute
Paired
Percent
CGM-
Differences
YSI
3023
16.7%
680
10.7%
3108
11.4%
777
8.0%
3021
11.9%
806
8.5%
Median
Percent
Absolute
Within
Percent
15/15%
Differences
YSI
13.2%
59%
7.9%
77%
8.2%
77%
6.4%
89%
8.9%
76%
7.2%
90%
Original
1
) - (Adult)
Percent
Percent
Within
Within
20/20%
30/30%
YSI
YSI
71%
86%
84%
96%
87%
95%
96%
99%
87%
95%
97%
99%
(SW10050) or
Percent
Greater
than
40/40%
YSI
6%
2%
2%
0%
2%
0%
292