Table 12-B. System Difference to SMBG Within CGM Glucose
Ranges (Pediatric)
CGM
Glucose
Study
1
Range
(mg/dL)
Original
Overall
Software 505
Original
*40-60
Software 505
Original
*61-80
Software 505
Original
81-180
Software 505
Original
181-300
Software 505
Original
301-350
Software 505
Original
351-400
Software 505
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
For CGM ≤ 80 mg/dL, the differences in mg/dL are included instead of percent
*
differences (%).
Dexcom G5 Mobile System User Guide
Technical Information
Number
2
of Paired
CGM-SMBG
16,318
4264
487
240
1340
399
7084
1650
5627
1526
1176
319
604
130
Mean
Median
Percent
Percent
Difference
Difference
2.2%
0.9%
-0.7%
-1.1%
-22.1
-17.0
-15.9
-14.0
-11.8
-8.0
-7.8
-6.0
1.1%
-1.0%
-1.2%
-2.6%
5.7%
3.4%
1.7%
0.9%
9.6%
7.2%
6.7%
5.9%
12.7%
10.2%
12.0%
8.9%
Mean
Absolute
Percent
Difference
15.3%
12.5%
23.9
16.9
17.0
13.7
15.4%
12.1%
13.5%
10.1%
14.2%
11.8%
16.1%
15.7%
Original
(SW10050) or
Median
Absolute
Percent
Difference
11.1%
9.5%
18.0
14.0
11.0
10.0
11.4%
9.5%
9.5%
7.7%
10.4%
8.9%
11.9%
10.6%