Table 10-B. System Readings Within Wear Days (Pediatric)
Statistic
Study
Original
Mean
Software 505
Original
Median
Software 505
Original
Standard
Deviation
Software 505
1
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
2
A total of 108 sensors were included with the
included with the
Number of Readings Provided
The system is capable of providing a reading up to every 5 minutes, or up to
288 readings per day. For a variety of reasons, the system may not display a
glucose reading and readings are "skipped." Table 9-A and 9-B estimate the
number of readings you can expect to receive from the system over the entire
7-day period after calibration. Table 10-A and 10-B show the number of
readings you can expect to receive from the system within each system wear
day.
For the
Software 505 Adult
between 1497 and 1992 valid glucose readings (or more than 75% of the
expected number of readings) as seen in Table 9-A. Adjusted within each
system wear-day, the system in the
average of 98% of all expected glucose readings (288) as seen in Table 10-A.
Dexcom G5 Mobile System User Guide
Technical Information
1
Day 1
Day 2
97%
96%
99%
99%
6%
9%
Software 505
Study.
Study (SW10505), 96% of systems provided
Day 3
Day 4
96%
96%
95%
96%
95%
96%
99%
99%
99%
98%
99%
99%
10%
9%
12%
6%
12%
10%
Original
Software 505 Adult
Day 5
Day 6
94%
94%
93%
95%
99%
99%
97%
97%
14%
14%
15%
7%
Original
(SW10050) or
Study and 77 sensors were
Study provided an
All
Day 7
2
Days
92%
95%
93%
95%
98%
99%
98%
98%
17%
12%
12%
11%
299
Need help?
Do you have a question about the G5 and is the answer not in the manual?
Questions and answers