Table 6-B. Hyperglycemia Alert and Detection Rate Evaluation in Reference
to YSI 15 Minutes Before and After (Pediatric, Ages 6-17 Years)
Hyperglycemia
Alert Level
(mg/dL)
120
140
180
200
220
240
300
1
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
True
1
Study
Alert
Rate
Original
91%
Software 505
98%
Original
87%
Software 505
97%
Original
75%
Software 505
94%
Original
71%
Software 505
94%
Original
67%
Software 505
93%
Original
62%
Software 505
88%
Original
43%
Software 505
69%
False
Hyperglycemia
Alert
Detection Rate
Rate
9%
98%
2%
99%
13%
99%
3%
98%
25%
99%
6%
98%
29%
98%
6%
97%
33%
97%
7%
96%
38%
96%
12%
94%
57%
93%
31%
84%
Original
Hyperglycemia
Missed
Detection Rate
2%
1%
1%
2%
1%
2%
2%
3%
3%
4%
4%
6%
7%
16%
(SW10050) or
286