Dexcom G5 User Manual page 285

Mobile continuous glucose monitoring
Hide thumbs Also See for G5:
Table of Contents

Advertisement

Table 6-A. Hyperglycemia Alert and Detection Rate Evaluation in Reference
to YSI 15 Minutes Before and After (Adult)
Hyperglycemia
Alert Level
(mg/dL)
120
140
180
200
220
240
300
1
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
True
1
Study
Alert
Rate
Original
95%
Software 505
98%
Original
94%
Software 505
97%
Original
92%
Software 505
97%
Original
92%
Software 505
96%
Original
91%
Software 505
94%
Original
91%
Software 505
93%
Original
82%
Software 505
86%
False
Hyperglycemia
Alert
Detection Rate
Rate
5%
98%
2%
100%
6%
97%
3%
99%
8%
97%
3%
99%
8%
97%
4%
98%
9%
95%
6%
98%
9%
94%
7%
95%
18%
86%
14%
90%
Original
Hyperglycemia
Missed
Detection Rate
2%
0%
3%
1%
3%
1%
3%
2%
5%
2%
6%
5%
14%
10%
(SW10050) or
285

Advertisement

Table of Contents

Troubleshooting

loading
Need help?

Need help?

Do you have a question about the G5 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents