Table 2-B. Number and Percentage of YSI Values When CGM Readings Are
"LOW" or "HIGH" (Pediatric)
CGM
Study
Readings
Original
"LOW"
Software
505
CGM
Study
Readings
Original
"HIGH"
Software
505
1
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Agreement When CGM Reads "LOW" or "HIGH"
The system reports glucose readings between 40 and 400 mg/dL. When the
system determines the glucose reading is below 40 mg/dL, it displays "LOW"
in the receiver status box. When the system determines that the glucose level
is above 400 mg/dL, it displays "HIGH" in the receiver Status Box. Because
the system does not display glucose values below 40 mg/dL or above 400
Dexcom G5 Mobile System User Guide
Technical Information
CGM-YSI
1
Pairs
n
Cumulative
Percent
n
Cumulative
Percent
CGM-YSI
1
Pairs
n
Cumulative
Percent
n
Cumulative
Percent
YSI mg/dL
< 55
< 60
0
0
0%
0%
3
5
19%
31%
YSI mg/dL
> 340
> 320
> 280
38
51
54%
73%
14
19
58%
79%
< 70
< 80
≥ 80
0
0
0%
0%
100%
10
15
63%
94%
> 240
≤ 240
68
69
97%
99%
22
23
92%
96%
Original
(SW10050) or
Total
13
13
1
16
6%
Total
1
70
1%
1
24
4%
271