Table 11-B. System Agreement to SMBG Within CGM Glucose Ranges
(Pediatric)
CGM
Glucose
Study
1
Range
(mg/dL)
Original
Overall
Software 505
Original
40-60
Software 505
Original
61-80
Software 505
Original
81-180
Software 505
Original
181-300
Software 505
Original
301-350
Software 505
Original
351-400
Software 505
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number
2
of Paired
CGM-SMBG
16,318
4264
487
240
1340
399
7084
1650
5627
1526
1176
319
604
130
Percent
Percent
Within
Within
15/15%
20/20%
SMBG
SMBG
64%
76%
73%
84%
44%
55%
54%
71%
59%
70%
64%
76%
62%
74%
72%
84%
69%
80%
79%
89%
65%
77%
72%
83%
58%
72%
69%
79%
Percent
Within
30/30%
SMBG
89%
94%
68%
86%
85%
92%
90%
95%
90%
97%
90%
94%
86%
86%
Original
(SW10050) or
Percent
Greater
than
40/40%
SMBG
5%
2%
19%
7%
7%
2%
5%
2%
5%
2%
4%
2%
6%
8%