Table 1-A. System Agreement to YSI Within CGM Glucose Ranges (Adult)
CGM
Glucose
1
Range
(mg/dL)
Overall
Software 505
40-60
Software 505
61-80
Software 505
81-180
Software 505
181-300
Software 505
301-350
Software 505
351-400
Software 505
1
CGM readings are within 40-400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number of
2
Study
Paired
CGM-YSI
Original
9152
2263
Original
512
120
Original
781
226
Original
3853
738
Original
2784
798
Original
775
229
Original
447
152
Percent
Percent
Within
Within
15/15%
20/20%
YSI
YSI
71%
82%
86%
93%
67%
78%
89%
94%
73%
85%
91%
96%
67%
78%
84%
92%
72%
84%
86%
93%
82%
91%
86%
94%
74%
84%
80%
92%
Original
Percent
Percent
Greater
Within
than
30/30%
40/40%
YSI
YSI
92%
3%
98%
1%
88%
6%
98%
0%
94%
2%
99%
0%
91%
3%
98%
1%
93%
4%
98%
1%
97%
2%
98%
1%
91%
5%
97%
0%
(SW10050) or
267
Need help?
Do you have a question about the G5 and is the answer not in the manual?
Questions and answers