Dexcom G5 User Manual page 296

Mobile continuous glucose monitoring
Hide thumbs Also See for G5:
Table of Contents

Advertisement

In the
Original Adult
(94%) of the sensors lasted until Day 7 (145-168 hours). There were 6 (6%)
sensors that ended early, four of which lasted more than 3 days.
In the
Software 505 Adult
percent (98%) of the sensors lasted until Day 7 (145-168 hours). There was 1
(2%) sensor that ended early, which lasted until day 5 of the sensor wear.
In the
Original Pediatric
percent (85%) of the sensors lasted until Day 7 (145-168 hours).
In the
Software 505 Pediatric
percent (94%) of the sensors lasted until Day 7 (145-168 hours).
Table 9-A. Number of Readings Provided by Each Sensor Over 7 Days
(Adult)
% of Total Possible
Readings Provided
0-25%
26-50%
51-75%
76-100%
1
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Study, 108 sensors were evaluated. Ninety-four percent
Study, 51 sensors were evaluated. Ninety-eight
Study, 351 sensors were evaluated. Eighty-five
Study, 77 sensors were evaluated. Ninety-four
1
Study
Original
Software 505
Original
Software 505
Original
Software 505
Original
Software 505
Total Readings
Provided
(Min-Max)
167-491
0
719-914
856-856
1267-1267
1253-1253
1811-1992
1497-1992
Original
% of Systems Providing
That Number of Readings
2%
0%
4%
2%
1%
2%
94%
96%
(SW10050) or