Table 2-A. Number and Percentage of YSI Values When CGM Readings Are
"LOW" or "HIGH" (Adult)
CGM
Study
Readings
Original
"LOW"
Software
505
CGM
Study
Readings
Original
"HIGH"
Software
505
1
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
CGM-YSI
1
Pairs
n
Cumulative
Percent
n
Cumulative
Percent
CGM-YSI
1
Pairs
n
Cumulative
Percent
n
Cumulative
Percent
YSI mg/dL
< 55
< 60
66
84
42%
54%
11
16
61%
89%
YSI mg/dL
> 340
> 320
> 280
189
220
76%
89%
40
43
89%
96%
100%
< 70
< 80
≥ 80
123
142
79%
92%
17
18
94%
100%
> 240
≤ 240
238
246
96%
99%
45
45
100%
Original
(SW10050) or
Total
13
155
8%
0
18
0%
Total
2
248
1%
0
45
0%
270