Table 5-A. Hypoglycemia Alert and Detection Rate Evaluation in Reference to
YSI 15 Minutes Before and After (Adult)
Hypoglycemia
Alert Level
(mg/dL)
55
60
70
80
90
1
Both sets of study data are presented and are labeled as
Software 505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
True
1
Study
Alert
Rate
Original
50%
Software 505
71%
Original
64%
Software 505
85%
Original
79%
Software 505
92%
Original
87%
Software 505
95%
Original
90%
Software 505
96%
False
Hypoglycemia
Alert
Detection Rate
Rate
50%
71%
29%
68%
36%
75%
15%
83%
21%
83%
8%
91%
13%
86%
5%
90%
10%
89%
4%
94%
Original
Hypoglycemia
Missed
Detection Rate
29%
32%
25%
17%
17%
9%
14%
10%
11%
6%
(SW10050) or