Triggering the Oscilloscope
8.7.8

Runt Trigger

The runt trigger sets the oscilloscope to trigger pulses that cross one trigger level but
not another, as shown in the figure below.
Upper Limit of the
Trigger Level
Lower Limit of the
Trigger Level
Trigger Type
Click or tap the drop-down button of Type to select "Runt".
After a trigger type is selected, the current trigger setting information (including
trigger type, trigger source, and trigger level) is displayed in the trigger information
label at the top of the screen, as shown in the figure below. The information will
change based on the trigger settings.
HDO1000 User Guide
86
Positive Runt Pulse

Figure 8.15 Runt Trigger

Figure 8.16 Runt Trigger Setting Menu

Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Negative Runt Pulse