VTI Instruments Metek EX1202 User Manual page 74

High-density switch/measure and control systems
Hide thumbs Also See for Metek EX1202:
Table of Contents

Advertisement

VTI Instruments Corp.
Sample Trigger Options
Sample Trigger Source: sets the trigger line that moves the DMM from the Wait-For-
Sample Trigger Slope: indicates whether a rising or falling edge will move the DMM from
Count: sets the number of samples to that will be taken per trigger event.
Delay: indicates the delay that will occur between getting a Sample Trigger to taking a
Interval: when set to Interval, an internal time periodically sends a sample trigger. This
Trigger Output Options (a trigger can be generated either when the measurement completes,
or when the read value is out of tolerance)
Measurement Complete Trigger: sets the trigger line that will be pulsed when the
Over Limit Trigger: sets the trigger line to pulse when an over-limit condition occurs.
High Limit: sets the high input level for the over-limit condition.
Low Limit: sets the low input level for the over-limit condition.
74
Sample state to the Take Measurement state.
the Wait-For-Sample state to the Take Measurement state.
measurement.
parameter sets the time between sample triggers.
Measurement Complete state is reached.
Basic EX1200 DMM Operation