When enabled as an output, each channel also has the ability to generate a 1 µs pulse upon
command. An example application of this pulse is to use the EX1200 to externally trigger another
piece of test equipment. The specific operation of the pulse depends on the static level
programmed for that channel. When a channel is programmed with a static level of high, the pulse
will be low-going. When a channel is programmed with a static level of low, the pulse will be
high-going. Each pulse generation requires a separate function call.
For expanded and more automated operation, each DIO channel can be independently linked to
any of the EX1200's backplane trigger lines, allowing any trigger condition (such as an over-limit,
or measurement complete event) that can be asserted by a plugin-module or the integrated DMM
on those backplane lines to be exported to any of the DIO channels.
The electrical specifications for the digital I/O port are provided in Table 3-. Particular note should
be given to the V
voltage source, even through a resistance, could permanently damage the EX1200.
Trigger Bus
The EX1200 provides an compatible trigger bus connector set on the rear panel of the instrument.
Both connectors have the same pin assignments and are wired in parallel. For more information on
the Trigger Bus and creating an appropriate cable, please visit
LXI Standard Revision 1.2 and LXI Trigger Bus Cable and Terminator Specifications Rev 1.1.
For proper operation, each trigger bus connector must be populated with either a cable or a
termination. Since the two connectors per unit are in parallel, they are electrically equivalent with
respect to receiving the cable or termination. VTI Instruments P/N: 70-0322-000 provides two
64
specification of -0.5 V to 5.5 V. Exceeding that value with an external
INPUT
Characteristic
V
INPUT
V
IH
V
IL
V
(I
= -32 mA)
OH
OH
V
(I
= 64 mA)
OL
OL
T
3-2: D
I/O P
ABLE
IGITAL
Pin
Function
1
+3.3 V
2
GND
RP_TRIG_P1
3
4
RP_TRIG_N1
5
GND
RP_TRIG_P3
6
7
RP_TRIG_N3
8
GND
RP_TRIG_P5
9
RP_TRIG_N5
10
11
RESERVED
12
RP_TRIG_P7
RP_TRIG_N7
13
1
2
3
4
14
15
16
17
T
3-3: T
B
ABLE
RIGGER
US
Value
-0.5 V to 5.5 V
2 V min
0.8 V max
2 V min
0.55 V max
E
S
ORT
LECTRICAL
PECIFICATIONS
www.lxistandard.org
Pin
Function
14
RP_TRIG_P0
15
RP_TRIG_N0
RESERVED
16
17
RP_TRIG_P2
18
RP_TRIG_N2
GND
19
20
RP_TRIG_P4
21
RP_TRIG_N4
GND
22
RP_TRIG_P6
23
24
RP_TRIG_N6
25
RESERVED
5
6
7
8
9
10
11
12
13
18
19
20
21
22
23
24
25
C
P
A
ONNECTOR
IN
SSIGNMENT
EX1200 Series Basic Mainframe Operation
and refer to