Figure 9
Indirect connection
JD785A/JD786A
CAUTION
The maximum power for the Spectrum Analyzer RF In port is +25 dBm (0.316 W). If the
input signal level to be measured is greater than +25 dBm, use a High Power
Attenuator to prevent damage when you directly connect the signal to the instrument or
connect the signal from the coupling port of a directional coupler.

Selecting measurement mode

Procedure
1.
Press the MODE hard key.
2.
Press the Spectrum Analyzer soft key. The Measure Off mode is selected by default.
3.
To select a measurement mode, press the MEASURE hot key and then select the measurement
mode:
Channel Power
Occupied BW
Spectrum Emission Mask
Adjacent Channel Power
Multi-ACP
Spurious Emissions
AM/FM Audio Demod
Field Strength
Route Map
Dual Spectrum
PIM Detection > Single Carrier or Multiple Carrier

Configuring test parameters

Setting frequency

You need to set the frequency range to be measured with either the center frequency/span or the
start/stop frequencies. You can also set the frequency with the channel number and span.
Using Spectrum Analyzer
JD780A Series Analyzers
JD788A
59