JD780A Series Analyzers
4.
Optional. Press the More (1/2) > Trace Info soft key, and then select the trace number to view
the trace's parameter setting information stored at the time of the measurement or None to hide
the information display.
5.
Optional. If you have the two traces T1 and T2, you can perform trace math. To view the power
difference between the traces, press the T1 – T2 -> T5 or T2 – T1 -> T6 soft key.
The result is overlaid on the screen along with the second Y-axis.
Conducting RF measurements
Channel power
The Channel Power measurement is a common test used in the wireless industry to measure the total
transmitted power of a radio within a defined frequency channel. It acquires a number of points
representing the input signal in the time domain, transforms this information into the frequency domain
using Fast Fourier Transform (FFT), and then calculates the channel power. The effective resolution
bandwidth of the frequency domain trace is proportional to the number of points acquired for the FFT.
The channel power measurement identifies the total RF power and power spectral density (PSD) of
the signal in the LTE channel bandwidth.
Setting measure setup
After configuring test parameters as described in the "Configuring test parameters" on page 197, you
can continue your measurement.
Procedure
1.
Press the MEASURE SETUP hot key.
2.
Press the Bandwidth soft key.
3.
Select the nominal channel bandwidth to be measured, from the following choices:
–
1.4 MHz, 3 MHz, 5 MHz, 10 MHz, 15 MHz, and 20 MHz
204
Using LTE - FDD Signal Analyzer