Press the Slot No soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the Enter soft key.
4.
Toggle the Cell ID soft key and select Auto or Manual as desired:
Auto: to let the instrument detect the Cell ID for the LTE signal automatically.
Manual: to set a specific Cell ID for the LTE signal manually in order to speed up the
synchronization with a BTS.
5.
Optional. Press the Miscellaneous soft key, and then do the following(s) as needed:
To select the number of antenna ports, toggle the MIMO soft key and select 2x2 or 4x4.
This MIMO (Multiple Input Multiple Output) setting is activated only if the option 031 LTE-
Advanced TDD is installed with a license number. If not, the instrument sets this option to
2x2 by default. A 2x1 or 4x1 RF combiner is also required to able to test on MIMO
channels.
To assign a antenna port number automatically or manually, press the Antenna Port soft
key and select the option: Auto, 0, 1, 2, and 3. If the option 031 LTE-Advanced TDD is not
installed or if the MIMO is set to 2x2, the antenna ports 2 and 3 are disabled.
To select the cyclic prefix, toggle the Cyclic Prefix soft key and select Normal or
Extended. See "Appendix E – Cyclic prefix table" on page 575 for OFDM parameter details
by different bandwidth.
To set the delay, complete the following steps:
a.
b.
c.
NOTE
The Delay setting is used only when there is a time offset in the signals to be measured.
Setting limit for power vs. time tests
Procedure
1.
Press the LIMIT hot key.
2.
Press the PvsT Test Limits soft key.
3.
Select the test item(s) and set the limit(s):
To set the limit for
Slot average power
Off power
Transition period
4.
Optional. You can enable alarm sound that goes off if the measurement falls outside of the limit.
Toggle the Beep soft key between On and Off to enable or disable the beep sound.
5.
Optional. Go to SAVE/LOAD > Save, and then select Limit to save the limit settings.
See "Using save" on page 31 for more information.
Press the Delay soft key to set the amount of delay in µs.
Enter a value by using the numeric keys. You can also use the rotary knob.
Press the µs soft key.
Select
Slot Avg Power
Off Power
Transition Period
Using LTE – TDD Signal Analyzer
JD780A Series Analyzers
Set
High Limit,
Low Limit
High Limit
High Limit