Connect the Reflection/RF Out port and the RF In port for the Cable and Antenna Analyzer with
the test cable and the proper adapter as illustrated in Figure 248-2.
10.
Press the Continue soft key to continue calibration.
The calibration progress bar appears.
After completion, the calibration status on the screen changes to ON.
O-S-L-Thru calibration using electronic EZ-Cal kit for vector measurements
Figure 249 illustrates the connection using the EZ-Cal kit for calibration in Vector measurements.
Figure 249 2-port calibration connection with EZ-Cal in vector measurements
1)
Procedure
1.
Connect an extension cable, if necessary, to the Reflection/RF Out port of the instrument.
2.
Connect your EZ-Cal to the other open end of the extension cable or to your instrument's
Reflection/RF Out port directly.
3.
Connect the USB cable to the side of the EZ-Cal and then to the USB Host port to power the
e-calibration kit as illustrated in Figure 249-1.
The instrument detects the EZ-Cal automatically and displays the EZ-Cal icon.
4.
Press the MEASURE SETUP hot key.
5.
Press the Calibrate soft key.
The on-screen instruction appears to guide you through the e-calibration.
6.
Connect the Reflection/RF Out port and the RF In port for the Cable and Antenna Analyzer with
the test cable and the proper adapter as illustrated in Figure 249-2.
7.
Press the Continue soft key to continue calibration.
The calibration progress bar appears.
After completion, the calibration status on the screen changes to ON.
Scalar measurement
Figure 250 illustrates the connection for calibration in Scalar measurements.
498
2)
Using Cable and Antenna Analyzer