Calibration; Offset (20,9%) Calibration - Ametek mocon Dansensor CheckMate3 User Manual

Table of Contents

Advertisement

46
User Guide

Calibration

NOTE! Apart from the below described "Offset (20,9%) calibration" all calibration
procedures must be carried out by Service Technicians only, as calibration requires
for special skills and the access to certified calibration gasses.
Invalid calibration compromises measurement results.

Offset (20,9%) calibration

NOTE! Only applies to O
Follow procedure below to perform the calibration:
1. From the measuring screen push the key.
2. If you are not already in Supervisor mode select Access
level from the main menu.
3. To enter the pin code push the or
4. The standard code from the factory is "0", but it may have
been changed.
Use the appearing touch-screen keyboard to enter the
appropriate code and push "OK".
5. In the Main menu use
Calibration menu item then push the key to enter the
function.
P/N 300042-I
05/2019
EN
EC (Electro-chemical) sensors.
2
key.
key to scroll down to the
®
Dansensor
CheckMate 3
Product: 47
08/17/06 13:46
Meat 400 gr.
Measuring mode: Manual spot
Product
Menu
0,482
29,5
%
%
O
CO
2
2
70,0
Bal
Press
to measure
13:46
Access level
08/17/06
lock instrument in user level
Current access level: User
Enter pincode:
Enter access pin code:
*
1
2
3
+/-
4
5
6
Del
7
8
9
CA
0
00 000 OK
Main menu
13:46
08/17/06
Product menu
Data log
Diagnostic menu
General setup
Access level
Calibration
****

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents