Gas Connection; Basic Specifications - Ametek mocon Dansensor CheckMate3 User Manual

Table of Contents

Advertisement

52
User Guide

Gas connection

Gas inlet
Gas outlet

Basic specifications

Warm-up time
Measuring ranges
Ambient temperature
Ambient pressure
Measurement gas
Measurement gas flow
Measurement gas temperature
Sample pressure range
1
Warm-up time is shortened at temporary power interrupts.
2
The flow will be lower than 27 ml/min when sample pressure is lower than atmospheric pressure.
P/N 300042-I
05/2019
EN
Dansensor
Sample hose:
Sample type/connector:
Filter/Water trap:
Connection for hose (inner diameter Ø 3mm)
Operational after 10 min.
(60 minutes to full specifications)
0-100 % O
and 0-100% CO
2
Operational:
+0 to +45 °C, less than 95 %RH,
non condensing
Storage:
-20°C to +60°C, less than 95 %RH,
non condensing
Operational:900 hPa to 1050 hPa
Inert gasses (O
, CO
, N
2
2
2
non condensing.
27 to 40 ml/min. (Typical flow 34 ml/min.)
5 °C to 35 °C
Ambient +100 mbar -100mbar
®
CheckMate 3
Length 60 cm, Ø0,5mm
Needle 0,8x40mm
or
SmartPen 0,8x9mm
0,2 μ
1
2
or Ar), less than 95 %RH,
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents