Calibration; Auto Calibration - Ametek ta7000R User Manual

Trace analytical gas purity monitor
Hide thumbs Also See for ta7000R:
Table of Contents

Advertisement

Calibration Page
Pressing the Page key three times will display the calibration informa-
tion page.
The calibration screen shows information about the most recent calibra-
tion run.
Name. is the gas species to be measured.
Conc is the cylinder concentration (or blend) of the calibration gas
connected to the instrument.
Area is the area of the last analysis run for each gas type.
Rfactor is the response factor created by the most recent (or previ
ous) calibration cycle. Rfactor is used to generate the area.
RT. is retention time. This is the time point of the peak from the
start of the most recent analysis run.
Conc Factor is a figure that is preset at the factory. This number
represents the internal gas blender ratio.
Calibration Screen

Auto Calibration

The Calibration screen is also used to setup automatic calibrations. The
analyzer does not need zero calibration. An external calibration gas
source must be connected to the appropriate port on the rear panel of
the analyzer marked AUX. Calibration gas specifications are located in
the specifications pages of this manual.
Idle Mode

Calibration

Name
Conc.
Area
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
8100
H2
*
7500
CO
Conc. Factor 200
Cal Time
00:00
Cal Purge
1 Min
12:34:25
Rfactor
RT
1346165
11737367
Interval
None
Cal Tol
10%
Edit
Page
User Manual | 39