Data Log; Diagnostic Menu - Ametek mocon Dansensor CheckMate3 User Manual

Table of Contents

Advertisement

36
User Guide

Data log

Selecting Data log from the Main menu will display a screen with a list of all the logged data
for the currently selected product.
Measuments which have been interrupted or marked as invalid are displayed on a red
background.
Data log
Product Meat
:
Mix 400gr.
Data collection: 21
Sample time
28/8/05 16:46:18
28/8/05 16:46:10
28/8/05 16:45:59
28/8/05 16:45:51
28/8/05 16:45:45
28/8/05 16:45:37
From here the logged data can be printed by pressing the
printer only).
Logged data can be exported (see
"Product menu" on page 33
Data log memory full
If data log memory is full an error will occur (Error code: 2679, No free space on flash).
In this case you must empty data log by using the functions for deleting of data collections -
see
"Product menu" on page 33
Please note that new measurements will not be logged until current data logs have been
deleted.

Diagnostic menu

Selecting Diagnostic menu from the Main menu will display a screen showing the internal
parameters of the device.
Diagnostic
Current user
Device temperature
Atmospheric pressure
Sample hose
Package pressure
O sensor (EC)
2
These are values such as current user, device's internal temperature, the pressure in the
sensors, the sensor's conditions, serial numbers for main components, day/hour counters,
time to calibration, and software versions, etc.
Pressing the
product.
The parameters can only be read and not changed.
P/N 300042-I
05/2019
13:46
08/17/06
%CO2
Bal
mbar
XX.X Low XX.X XXX.X
XX.X
XX.X XXX.X
XX.X
XX.X XXX.X
XX.X Low XX.X XXX.X
XX.X
XX.X XXX.X
XX.X Low XX.X XXX.X
"Export/Import" on page 49
for details).
for details.
13:46
08/17/06
USER1
XX,X °C
XXXX,X mbar
XXX,X mbar
XXX,X mbar
key brings up an Error log screen with a list of all errors for the selected
EN
Printing
13:46
08/17/06
Data collections
Statistics
key (devices with built-in
for details) or deleted (see
Diagnostic
Error log
28/8/2005 16:45 Er. xxxx
Device temperature to high
26/8/2005 11:45 Er. xxxx
O2 sensor temperature to high
6/8/2005 10:45 Er. xxxx
xxxxxx
6/8/2005 10:44 Er. xxxx
xxxxxx
6/8/2005 10:35 Er. xxxx
xx xxxxxx xx xx xx xxxxxx
6/8/2005 10:01 Er. xxxx
®
Dansensor