Ametek mocon Dansensor CheckMate3 User Manual page 43

Table of Contents

Advertisement

®
Dansensor
CheckMate 3
Start
Column
Length ** Type
pos.*
1
1
8
Numeric
2
10
8
Numeric
3
19
8
Numeric
4
28
8
Numeric
5
37
1
Numeric
6
39
1
Numeric
7
41
1
Numeric
8
43
6
Numeric
9
50
8
Text
10
59
8
Text
11
68
40
Text
12
109
40
Text
13
150
24
Text
14
175
40
Text
15
216
37
Text
16
254
37
Text
17
292
37
Text
18
330
37
Text
19
368
37
Text
20
406
100
Text
21
507
14
Text
22
522
1
Numeric
23
524
6
Numeric
24
531
6
Numeric
25
538
1
Numeric
26
540
1
Numeric
27
542
8
Numeric
28
551
1
Numeric
29
553
1
Numeric
30
555
8
Numeric
31
564
1
Numeric
32
566
1
Numeric
33
568
8
Numeric
34
577
1
Numeric
35
579
1
Numeric
36
581
8
Numeric
37
590
1
Numeric
38
592
1
Numeric
39
594
8
Numeric
40
603
1
Numeric
41
605
1
Numeric
42
607
8
Numeric
43
616
1
Numeric
44
618
1
Numeric
45
620
1
Numeric
46
622
1
Numeric
47
624
1
Numeric
48
626
1
Numeric
49
628
6
Numeric
50
635
6
Numeric
51
642
1
Numeric
Total:
644
Chars in ASCII (x4 in UNICODE) +CRLF
* Only when all elements are selected and format is ASCII.
** Length: Fixed length. When unicode length must be multiplied by 4.
Fixed decimal placement. Leading zero in numbers. Gas concentrations are always in % buth with ppm resolusion (regardless of accuracy) e.g. "020.9465"
Example:
000,8395;000,3200;098,8405;000,0000;0;0;0;000002;27/02/08;16:43:39;Product 2;
EN
Serial data dump specification
Description
O2 [%]
CO2 [%]
Balance [%]
(Not used)
O2 alarm [0=No alarm 1=High Alarm 2=Low Alarm]
CO2 alarm [0=No alarm 1=High Alarm 2=Low Alarm]
(Not used)
Product number [Integer]
Date [MM/DD/YY] / [DD/MM/YY]
Time [HH:MM:SS]
Product name [Text]
Product barcode [Text]
CheckMate II S/N [Text]
User ID [Text]
User field 1 [Text]
User field 2 [Text]
User field 3 [Text]
User field 4 [Text]
User field 5 [Text]
Note [Text]
SW version [Text]
Measure mode
[0=Continuous, 1=manual spot, 2=auto spot 3=intermitted]
Sample time [seconds]
Measure delay at intermitted [minutes]
Alarm 1 type [0=Off, 1=high, 2=Low]
Alarm 1 gas [0=O2, 1=CO2
Alarm 1 concentration [%]
Alarm 2 type[0=Off, 1=high, 2=Low]
Alarm 2 gas [0=O2, 1=CO2
Alarm 2 concentration [%]
Alarm 3 type[0=Off, 1=high, 2=Low]
Alarm 3 gas [0=O2, 1=CO2
Alarm 3 concentration [%]
Alarm 4 type[0=Off, 1=high, 2=Low]
Alarm 4 gas [0=O2, 1=CO2
Alarm 4 concentration [%]
Alarm 5 type [0=Off, 1=high, 2=Low]
Alarm 5 gas [0=O2, 1=CO2
Alarm 5 concentration [%]
Alarm 6 type[0=Off, 1=high, 2=Low]
Alarm 6 gas [0=O2, 1=CO2
Alarm 6 concentration [%]
Note [0=Off, 1=At alarm, 2=At no alarm, 3= Always, 4=Always]
User field 1 required [0=no, 1=Always, 2=Once]
User field 2 required [0=no, 1=Always, 2=Once]
User field 3 required [0=no, 1=Always, 2=Once]
User field 4 required [0=no, 1=Always, 2=Once]
User field 5 required [0=no, 1=Always, 2=Once]
Device temperature [°C]
Atmospheric pressure [mbar]
Invalid measurement [0=No, 1=Yes]
User Guide
Note
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
No decimal point (e.g. "000032")
Variable according to set-up
Left alligned
Fixed decimal point plus sign, eg. "+005.0"
No decimal point
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
Fixed decimal point, eg. "020.9465"
Signed value and fixed point
(eg. "+023.0" or "-003.2")
No decimal point
P/N 300042-I
05/2019
41