General Setup - Ametek mocon Dansensor CheckMate3 User Manual

Table of Contents

Advertisement

®
Dansensor
CheckMate 3

General setup

Selecting General setup from the Main menu will display a menu with available set-up
parameters.
General setup
Data log setup
...
Network setup ...
Custom fields setup ...
User setup ...
Formats/Units/Time...
COM1=
Press. Diff. Alarm high limit
Menu items followed by ... brings up a submenu with available set-up parameters for the
selected item.
The General setup menu holds the following items:
Data log setup ...
Network setup ...
Custom fields setup ...
User setup ...
Formats/Units/Time ...
COM1=
Press. Diff. Alarm high limit The settings control the window for the sample system's
Press. Diff. Alarm low limit
Back light
Contrast
Brightness
Measure delay
EN
13:46
08/17/06
Normal
75,0 mbar
Opens the Data log setup screen
See
"Data log setup" on page 38
Opens the Network setup screen
See
"Network setup" on page 42
Opens the Custom fields setup screen
See
"Custom fields" on page 42
Opens the User setup screen
See
"User setup" on page 43
Opens the Formats/Units/Time screen
See
"Formats/Units/Time" on page 44
Select intended use of the COM-port.
Normal
OEM
operating pressure. Default setting is 25.0 mbar (low limit) and
75.0 mbar (high limit).
Special applications may require adjustment of these values.
NOTE! As changing the values may impair
measurements, a warning is displayed when you
attempt to change the values.
Adjust display background light (1-5)
Adjust display contrast (1-10)
When display shows the measuring screen you can use the
and
Adjust display brightness (1-10)
Time before measurement starts after pressing the
(1.0 - 999.9 sec.)
for details.
for details.
for details.
for details.
for details.
Use as serial data dump port
Use for special applications
keys to adjust screen contrast.
User Guide
37
key.
P/N 300042-I
05/2019