6. The appearing Calibration menu holds a list of the
various sensors in the device.
For each sensor there is a table showing the appropriate
calibration parameters.
To proceed with the calibration push the key.
7. From this menu you can select between the following
items:
- Perform calibration
- Erase/enter calibration
Select Perform calibration, then push the key.
8. Use
or
key to select the calibration value fields for
the O
sensor (EC), then push the
2
9. The pump starts and now make sure to expose the
needle to a 20,946% O
Let measurement run for approx. 5 min. then select the
20,9460 gas entry from the list and push
10. You will be asked to perform the calibration.
NOTE! Existing calibration value, if any, will be
overwritten.
Select "Yes" then push
EN
key.
gas.
2
key.
key.
User Guide
13:46
Calibration
08/17/06
Calibrate
O sensor (EC)
2
O
Response
Pressure Calibration
2
%
mV
mbar
date
0,0000
0,170
1032,0
05/08/06
20,9460
18,208
1031,9
05/08/06
CO sensor (100%)
2
CO
Response
Pressure Calibration
2
%
mbar
date
100,00
0,533
1034,0
05/08/06
60,00
0,463
1034,0
05/08/06
Calibrate
08/17/06
!!! WARNING !!!
Calibration requires use of certified gases
Wrong calibration will cause invalid measurements
Perform calibration
Erase / enter calibration
13:46
Calibration
08/17/06
Select sensor and press
to begin calibration
O sensor (EC)
2
O
Response
Pressure Calibration
2
%
mV
mbar
date
0,0000
X,XXX
XXXX,X
XX XX XX
/
20 9460
,
XX XXX
,
XXXX X
,
XX XX XX
/
CO sensor (100%)
2
CO
Response
Pressure Calibration
2
%
mbar
date
13:46
Calibrating
08/17/06
Select gas and press
Sensor:
XX,XXX
%O
EMF:
XX,XXX
mV
Pressure:
XXX,X
mbar
1:
0,0000
2:
20,9460
13:46
Warning
08/17/06
Measured value:
XX XXX
,
Calibration gas
:
XX XXX
,
Perform calibration?
Select and press
No
Yes
P/N 300042-I
05/2019
47
13:46
/
/
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents