Proline Promag 53 PROFIBUS DP/PA
No.
Device status message
(local display)
472
P: >< BATCH
QUANTITY
$: # 472
473
P: PROGRESS NOTE
$: # 473
474
P: MAX. FLOW RATE
!: # 474
Endress+Hauser
PROFIBUS measured value status
0x00
BAD
Non specific
0x01
(uncertain
0x02
status)
0x80
GOOD
O.K.
0x00
BAD
Non specific
0x01
(uncertain
0x02
status)
Extended
diagnostic
Cause/remedy
message in the
PROFIBUS
master
O.K.
Batch Quantity
Cause:
Low
EPD calibration not possible as the fluid conductivity is
High
either too low or too high.
– Underbatching:
– Overbatching:
Remedy:
Underbatching:
1. Increase fixed correction quantity.
2. If the batching quantity changes, the minimum
Overbatching:
1. Reduce fixed correction quantity.
2. If the batching quantity changes, the maximum
O.K.
Progress Note
Cause:
End of filling process approaching.
The running filling process has exceeded the
predefined batch quantity point for the display warning
message.
Remedy:
No measures required (if necessary prepare to replace
container).
O.K.
Flow Rate
Cause:
Low
Maximum flow value entered is
High
exceeded.
Remedy:
Reduce the flow value.
Troubleshooting
The minimum quantity was not reached.
The maximum permitted batching quantity was
exceeded.
batching quantity must be adjusted.
batching quantity must be adjusted.
147
Need help?
Do you have a question about the PROFI BUS Proline Promag 53 PROFIBUS DP and is the answer not in the manual?