Proline Promag 53 PROFIBUS DP/PA
No.
Device status message
(local display)
359
S: PULSE RANGE
to
!: # 359 to 362
362
No. # 5xx → Application error
501
S: SW.-UPDATE ACT.
!: # 501
502
S: UP-/DOWNLO. ACT.
!: # 502
571
P: RUN BATCHING
!: # 571
572
P: HOLD BATCHING
!: # 572
Endress+Hauser
PROFIBUS measured value status
0x54
UNCERTAIN
Engineering
0x55
Unit Range
0x56
Violation
0x48
UNCERTAIN
Substitute set
0x49
(Substitute set
0x4A
of failsafe
status)
0x48
UNCERTAIN
Substitute set
0x49
(Substitute set
0x4A
of failsafe
status)
0x80
GOOD
O.K.
0x80
GOOD
O.K.
Extended
diagnostic
Cause/remedy
message in the
PROFIBUS
master
O.K.
Flow is Out of
Cause:
Low
Range
Pulse output: The pulse output frequency is out of
High
range.
Remedy:
1. Increase the setting for pulse weighting.
2. When selecting the pulse width, choose a
3. Reduce flow
O.K.
New amplifier
Cause:
Low
software loaded
New amplifier or communication software version
High
is loaded. Currently no other functions are possible.
Remedy:
Wait until process is finished. The device will
restart automatically.
O.K.
Upload/download
Cause:
Low
of device data
Up- or downloading the device data via
High
active
configuration program. Currently no other
functions are possible.
Remedy:
Wait until process is finished.
O.K.
Batch running
Cause:
Batching is started and active (valves are open)
Remedy:
No measures needed (during the batching process
some other functions may not be activated).
O.K.
Batch hold
Cause:
Batching has been interrupted (valves are closed)
Remedy:
Using either:
• PROFIBUS DP/PA
• Local display:
Troubleshooting
value that can still be processed by a connected
counter (e.g. mechanical counter, PLC, etc.).
Determine the pulse width:
– Version 1: Enter the minimum duration that
a pulse must be present at the connected
counter to ensure its registration.
– Variant 2: Enter the maximum (pulse)
frequency as the half "reciprocal value" that
a pulse must be present at the connected
counter to ensure its registration.
Example:
The maximum input frequency of the
connected counter is 10 Hz. The pulse width
to be entered is:
1
= 50 ms
2 . 10 Hz
– Continue batching with "GO ON".
– Interrupt batching with "STOP".
a0004437
143