Endress+Hauser Proline 300 Brief Operating Instructions
Endress+Hauser Proline 300 Brief Operating Instructions

Endress+Hauser Proline 300 Brief Operating Instructions

Measuring device for total solids measurement

Advertisement

Quick Links

KA01672D/06/EN/01.23-00
71637051
2024-03-15
Products
Brief Operating Instructions
Measuring device for total
solids measurement
Proline 300
HART transmitter
for total solids measurement via microwave
transmission
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Brief Operating Instructions part 2 of 2: Transmitter
Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →  3
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline 300 and is the answer not in the manual?

Questions and answers

Summary of Contents for Endress+Hauser Proline 300

  • Page 1 KA01672D/06/EN/01.23-00 71637051 2024-03-15 Brief Operating Instructions Measuring device for total solids measurement Proline 300 HART transmitter for total solids measurement via microwave transmission These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device.
  • Page 2 Measuring device for total solids measurement Proline 300 Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3 These Brief Operating Instructions are Brief Operating Instructions Part 2: Transmitter. The "Brief Operating Instructions Part 1: Sensor" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer •...
  • Page 4: Table Of Contents

    Table of contents Measuring device for total solids measurement Proline 300 Table of contents About this document ............. . 5 Symbols used .
  • Page 5: About This Document

    Measuring device for total solids measurement Proline 300 About this doc About this document Symbols used 1.1.1 Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury.
  • Page 6 About this document Measuring device for total solids measurement Proline 300 Symbol Meaning Potential equalization connection (PE: protective earth) Ground terminals that must be connected to ground prior to establishing any other connections. The ground terminals are located on the interior and exterior of the device: •...
  • Page 7: Safety Instructions

    Measuring device for total solids measurement Proline 300 Safety instru Safety instructions Requirements for the personnel The personnel must fulfill the following requirements for its tasks: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 8: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 9: Device-Specific It Security

    Measuring device for total solids measurement Proline 300 Safety instru Device-specific IT security The device offers a range of specific functions to support protective measures on the operator' s side. These functions can be configured by the user and guarantee greater in-operation safety if used correctly.
  • Page 10: Product Description

    Product description Measuring device for total solids measurement Proline 300 Product description The device consists of a transmitter and a sensor. The device is available as a compact version: The transmitter and sensor form a mechanical unit. Product design N ic...
  • Page 11: Mounting Procedure

    Measuring device for total solids measurement Proline 300 Mounting pro Mounting procedure Mounting the sensor For detailed information about mounting the sensor, see the Sensor Brief Operating Instructions →  3 Mounting the transmitter 4.2.1 Turning the transmitter housing To provide easier access to the connection compartment or display module, the transmitter housing can be turned.
  • Page 12 Mounting procedure Measuring device for total solids measurement Proline 300 4 mm 7 Nm (5.2 lbf ) A0043150  3 Ex housing Loosen the fixing screws. Turn the housing to the desired position. Tighten the securing screws. Endress+Hauser...
  • Page 13: Weather Protection Cover

    Measuring device for total solids measurement Proline 300 Mounting pro 4.2.2 Turning the display module The display module can be turned to optimize display readability and operability. – – – 3 mm A0030035 Depending on the device version: Loosen the securing clamp of the connection compartment cover.
  • Page 14: Transmitter Post-Installation Check

    Mounting procedure Measuring device for total solids measurement Proline 300 Transmitter post-installation check The post-installation check must always be performed after the following tasks: • Turning the transmitter housing • Turning the display module Is the device undamaged (visual inspection)? ...
  • Page 15: Electrical Connection

    Measuring device for total solids measurement Proline 300 Electrical conn Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 16 Electrical connection Measuring device for total solids measurement Proline 300 Pulse /frequency /switch output Standard installation cable is sufficient Relay output Standard installation cable is sufficient. Current input 0/4 to 20 mA Standard installation cable is sufficient Status input Standard installation cable is sufficient Cable diameter •...
  • Page 17: Connecting The Measuring Device

    Measuring device for total solids measurement Proline 300 Electrical conn Cable impedance (pair) Minimum 80 Ω Capacitance: core/shield Maximum 1 000 nF for Zone 1, Class I, Division 1 Maximum 24 µH/Ω for Zone 1, Class I, Division 1 5.2.3...
  • Page 18 Electrical connection Measuring device for total solids measurement Proline 300 5.3.1 Connecting the transmitter A0026781 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output or terminal connection for network connection via service interface (CDI-RJ45); optional: connection for external WLAN antenna or...
  • Page 19 Measuring device for total solids measurement Proline 300 Electrical conn N ic N ic te r te r e r g ö ff e r g ö ff iz e iz e te n te n r ir r ir...
  • Page 20 Electrical connection Measuring device for total solids measurement Proline 300 22 mm 24 mm – A0029816 10. Connect the cable according to the terminal assignment.  Signal cable terminal assignment: The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 21: Potential Equalization

    Measuring device for total solids measurement Proline 300 Electrical conn 5.3.2 Connecting the remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra. • The measuring device is always supplied with a dummy cover when the remote display and operating module DKX001 is ordered directly with the measuring device.
  • Page 22: Special Connection Instructions

    Electrical connection Measuring device for total solids measurement Proline 300 Special connection instructions 5.5.1 Connection examples Current output 4 to 20 mA HART 4...20 mA A0029055  5 Connection example for 4 to 20 mA HART current output (active) Automation system with current input (e.g. PLC) Ground cable shield at one end.
  • Page 23 Measuring device for total solids measurement Proline 300 Electrical conn HART input 4...20 mA A0028763  7 Connection example for HART input with a common negative (passive) Automation system with HART output (e.g. PLC) Active barrier for power supply (e.g. RN221N) Ground cable shield at one end.
  • Page 24 Electrical connection Measuring device for total solids measurement Proline 300 4...20 mA A0028759  9 Connection example for 4-20 mA current output (passive) Automation system with current input (e.g. PLC) Active barrier for power supply (e.g. RN221N) Analog display unit: observe maximum load...
  • Page 25 Measuring device for total solids measurement Proline 300 Electrical conn Switch output A0028760  11 Connection example for switch output (passive) Automation system with switch input (e.g. PLC with a 10 kΩ pull-up or pull-down resistor) Power supply Transmitter: observe input values...
  • Page 26 Electrical connection Measuring device for total solids measurement Proline 300 Current input A0028915  13 Connection example for 4 to 20 mA current input Power supply Terminal box External device (for reading in the flow rate value in order to calculate the load rate)
  • Page 27: Ensuring The Degree Of Protection

    Measuring device for total solids measurement Proline 300 Electrical conn Ensuring the degree of protection The measuring device fulfills all the requirements for the degree of protection IP66/67, Type 4X enclosure. To guarantee the degree of protection IP66/67, Type 4X enclosure, carry out the following steps after electrical connection: Check that the housing seals are clean and fitted correctly.
  • Page 28: Post-Connection Check

    Electrical connection Measuring device for total solids measurement Proline 300 Post-connection check Are cables or the device undamaged (visual inspection)?  Do the cables used comply with the requirements →  15?  Does the supply voltage match the specifications on the transmitter nameplate ? ...
  • Page 29: Operation Options

    Measuring device for total solids measurement Proline 300 Operation o Operation options Overview of operation options A0034513 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM)
  • Page 30: Structure And Function Of The Operating Menu

    Operation options Measuring device for total solids measurement Proline 300 Structure and function of the operating menu 6.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Operation task-oriented Setup Diagnostics Operating menu for experts Expert function-oriented A0014058-EN ...
  • Page 31: Access To Operating Menu Via Local Display

    Measuring device for total solids measurement Proline 300 Operation o Access to operating menu via local display X X X X X X 19.184 mA 12.5 X X X X X X Language 20.50 Ã English Deutsch Español Français Xxxxxxx...
  • Page 32 Operation options Measuring device for total solids measurement Proline 300 6.3.1 Operational display Explanatory symbols for the measured value Status area • Depends on the device version, e.g.: The following symbols appear in the status area of the • : Total solids operational display at the top right: •...
  • Page 33 Measuring device for total solids measurement Proline 300 Operation o Numeric editor Confirms selection. Moves the input position one position to the left. Exits the input without applying the changes. Inserts decimal separator at the cursor position. Inserts minus sign at the cursor position.
  • Page 34: Access To The Operating Menu Via The Operating Tool

    Operation options Measuring device for total solids measurement Proline 300 Operating key Meaning Escape key combination (press keys simultaneously) In menu, submenu • Pressing the key briefly: • Exits the current menu level and takes you to the next higher level.
  • Page 35: System Integration

    Measuring device for total solids measurement Proline 300 System integ System integration For detailed information on system integration, see the Operating Instructions for the device →  3 • Overview of device description files: • Current version data for the device •...
  • Page 36: Commissioning

    Commissioning Measuring device for total solids measurement Proline 300 Commissioning Installation and function check Before commissioning the device: ‣ Make sure that the post-installation and post-connection checks have been performed successfully. • "Post-mounting check" checklist →  14 • "Post-connection check" checklist →  28...
  • Page 37: Configuring The Measuring Device

    Measuring device for total solids measurement Proline 300 Commiss Configuring the measuring device The Setup menu with its submenus and various guided wizards is used for fast commissioning of the measuring device. They contain all the parameters required for configuration, such as for measurement or communication.
  • Page 38: Diagnostic Information

    Diagnostic information Measuring device for total solids measurement Proline 300 Diagnostic information Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. The message about remedial measures can be called up from the diagnostic message, and contains important information on the fault.
  • Page 40 *71637051* 71637051 www.addresses.endress.com...