Proline Promag 53 PROFIBUS DP/PA
No.
Device status message
(local display)
698
S: DEV. TEST ACT.
!: # 698
Endress+Hauser
PROFIBUS measured value status
0x60
UNCERTAIN
Simulated
0x61
Value
0x62
(manually
specified
value)
Extended
diagnostic
Cause/remedy
message in the
PROFIBUS
master
O.K.
Device test via
Cause:
Low
Fieldcheck active
The measuring device is being checked on site via
High
the test and simulation device.
Troubleshooting
145
Need help?