Switching the System On/Off
12.2.2
Switch-off Sequence
NOTICE
1
2
3
4
5
6
7
78
Damage to the instrument when not adhering to the switch-off sequence
The switch-off sequence must be followed! When not adhering to the
below listed switch-off sequence, the laser can be damaged.
Observe the user manuals for external lasers
Please refer to the information from the documents provided by the laser
manufacturer for the external lasers. Pay particular attention to the laser
manufacturer's notes!
Save the image data (Experiments -> Save all)
Close the LAS X software: On the menu bar, select File >
Exit.
Turn off the lasers in the supply unit with the key switch on
the front side of the compact supply unit. The emission
warning indicator goes out.
If you are using a HyD RLD, switch it off at its supply unit:
If your system has an infrared laser (MP) or an optical
parametric oscillator (OPO), switch off the NDD detection
unit:
Now, shut down the workstation.
Switch off the lasers on the front side of the compact supply
unit.