Switching the System On/Off
12.1.2
Switch-off Sequence
NOTICE
1
2
3
4
5
6
74
Damage to the instrument when not adhering to the switch-off sequence
The switch-off sequence must be followed! When not adhering to the
below listed switch-off sequence, the laser can be damaged.
Observe the user manuals for external lasers
Please refer to the information from the documents provided by the laser
manufacturer for the external lasers. Pay particular attention to the laser
manufacturer's notes!
Save the image data (
Close the LAS X software: On the menu bar, select File >
Exit.
Turn off the lasers in the supply unit with the key switch on
the main switch board of the flexible supply unit. The
emission warning indicator goes out.
If your system has a white light laser, check whether the
emission warning indicator is lit. If the emission warning
indicator on the front side of the white light laser is lit, press
the key below it. This causes all internal white light laser
components to shut off and the emission warning indicator
to go out.
Switch off the white light laser with the key switch on the
front of the white light laser.
If you are using a HyD RLD, switch it off at its supply unit.
Experiments
Save all
->
)