Ambient Conditions
8.4.2
System with compact supply unit
Supply voltage
Frequency
Power connection
Power supply for USA
8.4.3
External Lasers
36
100 V~ to 240 V~ ± 10%, grounded
50/60 Hz
Three separate electric circuits with
Power supply voltage: 100 V - 120 V, fuse protection: 20 A
or:
Power supply voltage: 200 V - 240 V, fuse protection: 12 - 16 A
(One for the compact supply unit's power supply and one for
supplying peripheral devices)
To prevent damage to the instrument, the two types of
voltage (200 to 240 V~ and 100 to 120 V~) must not be
combined. Depending on the system configuration, further
electrical circuits may be required.
Terminals of type NEMA 5-15
Observe the user manuals for external lasers
Please refer to the information from the documents provided by the laser
manufacturer for the external lasers. Pay particular attention to the laser
manufacturer's notes!