Ambient Conditions
8.2
Vibrations
Vibrations must be reduced to a minimum; for that reason, do not set up the system near any
of the following sources of vibration: shakers, ultracentrifuges, pumps, compressors,
refrigerators, vending machines, elevators.
Maximum tolerable vibrations:
Frequency range [5 Hz – 30 Hz]
Frequency range [> 30 Hz]
8.3
Room Dimensions
The room should be as large as possible so that multiple operators can find room around the
system. The free space to the right, left and rear of the system should be 60 cm. In front of
the system there must be 110 cm of free space so that the service technicians can access
the system as necessary.
WARNING
Also make sure that doors and passages are at least 100 cm wide in order to be able to
safely transport the system.
The following minimum size should be planned for the space:
STED System
8.4
Electrical Connection Requirements
WARNING
8.4.1
System with flexible supply unit
Supply voltage
Frequency
34
less than 12,5 μm/s root mean square
less than 12,5 μm/s root mean square
Fire or overheating of the system possible
Electrical peripheral devices must be placed at least 10 cm (4") away
from the wall and from flammable substances to prevent overheating or
fire of the system.
Minimum space requirements (length x depth)
440 x 320 cm (14'5" x 10'6")
Electric shock is possible when using ungrounded sockets
This system is designed for connection to grounded (earthed) outlets.
The grounding plug performs an important safety function. To avoid the
risk of electrical shock or damage to the instrument, do not disable this
feature. Operation without grounded sockets is not permitted.
100 V~ to 240 V~ ± 10%, grounded
50/60 Hz