Switching the System On/Off

The system comes with a flexible or compact supply unit. Observe the switch on/off
sequence that applies to your supply unit.
NOTICE
NOTICE
12.1

System with Flexible Supply Unit

The sequence below describes how the components of the system with a flexible supply
unit must be switched on and off.
12.1.1

Switch-on Sequence

1
2
Damage is possible when contacting the specimen stage
When using an inverted microscope, the illuminator arm must be swung
back before the system start and LAS X start because the stage can
damage the condenser, the objective or the specimen during
initialization.
Objective damage is possible when contacting the specimen stage
When using an upright microscope, the specimen stage must be moved
down before the system start and LAS X start because it could come into
contact with the objective during initialization and damage the objective.
Observe the user manuals for external lasers
Please refer to the information from the documents provided by the laser
manufacturer for the external lasers. Pay particular attention to the laser
manufacturer's notes!
If you are using an EL6000 fluorescence lamp, switch it on.
If you are using a HyD RLD, switch it on at its supply unit.
You can find additional information on the HyD RLD in
Chapter 12.3
Switching the System On/Off