Power connection
Power supply for USA
Four separate electric circuits with
Power supply voltage: 100 V - 120 V, fuse protection: 20 A
or:
Power supply voltage: 200 V - 240 V, fuse protection: 12 - 16 A
(Two for the supply unit's power supply and two for supplying
peripheral devices).
To prevent damage to the instrument, the two types of
voltage (200 to 240 V~ and 100 to 120 V~) must not be
combined. Depending on the system configuration, further
electrical circuits may be required.
Two terminals of the NEMA 5-20 type for the supply unit
Terminals of the NEMA 5-15 type for the remaining
instruments
Ambient Conditions