Advertisement

Quick Links

Instructions for use
SMARTmatic PROPHY S19 - 1.011.6740
SMARTmatic PROPHY S19 K - 1.011.6741

Advertisement

Table of Contents
loading

Summary of Contents for KaVo SMARTmatic PROPHY S19

  • Page 1 Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 Table of contents Table of contents 1 User instructions............................. 4 2 Safety ................................6 2.1 Infection hazard ............................. 2.2 Technical condition ..........................2.3 Accessories and combination with other equipment................
  • Page 4: User Instructions

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 1 User instructions 1 User instructions Dear User Congratulations on purchasing this KaVo quality product. By following the instructions below you will be able to work smoothly, economically and safely.
  • Page 5 Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 1 User instructions CAUTION In cases which – if not prevented – could lead to minor or moderate injury. NOTICE In cases which – if not prevented – could lead to material damage.
  • Page 6: Safety

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 2 Safety | 2.1 Infection hazard 2 Safety The instructions for use are a component of the product and must be read carefully prior to use and be accessible at all times.
  • Page 7: Accessories And Combination With Other Equipment

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 2 Safety | 2.3 Accessories and combination with other equipment ▶ Do not use motors for unauthorised purposes. 2.3 Accessories and combination with other equipment Use of un-authorised accessories or un-authorised modifications of the device could lead to injury.
  • Page 8 Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 2 Safety | 2.5 Service and repair ▶ KaVo recommends specifying in-house service intervals where the medical device is brought to a professional shop for cleaning, servicing and a function check.
  • Page 9: Description Of The Product

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 3 Description of the product | 3.1 Intended use 3 Description of the product SMARTmatic PROPHY S19 (Mat. No. 1.011.6740) SMARTmatic PROPHY S19 K (Mat. No. 1.011.6741) 3.1 Intended use...
  • Page 10: Technical Specifications

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 3 Description of the product | 3.2 Technical Specifications 3.2 Technical Specifications Drive speed max. 20,000 rpm Speed transmission Cooling air flow 5.5 to 9.5 Nl/min Prophylaxis heads with a Doriot connector can be inserted.
  • Page 11: Transportation And Storage Conditions

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 3 Description of the product | 3.3 Transportation and storage conditions 3.3 Transportation and storage conditions NOTICE Startup after refrigerated storage. Malfunction. ▶ Prior to startup, very cold products must be heated up to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
  • Page 12: Startup And Shut Down

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 4 Startup and shut down 4 Startup and shut down WARNING Hazard from non-sterile products. Infection hazard for care provider and patient. ▶ Before first use and after each use, sterilise the medical device.
  • Page 13: Operation

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 5 Operation | 5.1 Attaching the medical device 5 Operation 5.1 Attaching the medical device CAUTION Detachment of the medical device during treatment. Injury or property damage A medical device that is not properly locked in place can become disconnected from the motor coupling and fall off.
  • Page 14: Attach The Prophylaxis Head

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 5 Operation | 5.3 Attach the prophylaxis head 5.3 Attach the prophylaxis head Note All prophylaxis heads with a Doriot connection can be attached. The clamping system is designed for drive shafts working in accordance with DIN EN ISO 14457 for metallic and ceramic materials: Ø2.35 0/-0.016mm and plastic...
  • Page 15: Removing The Prophylaxis Head

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 5 Operation | 5.4 Removing the prophylaxis head 5.4 Removing the prophylaxis head ▶ Remove the prophylaxis head axially from the handpiece. CAUTION Hazard due to incorrectly stored instrument.
  • Page 16: Checking For Malfunctions And Troubleshooting

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions CAUTION Missing or damaged O-rings.
  • Page 17: Reprocessing Steps In Accordance With Iso 17664

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use WARNING Hazard from non-sterile products.
  • Page 18: Automated Internal Cleaning

    7 Reprocessing steps in accordance with ISO 17664 | 7.3 Disinfection ▶ Cover the medical device with the KaVo CLEANpac bag, and place it on the cor- responding care adapter. Press the spray button three times for 2 seconds each time.
  • Page 19: Manual Internal Disinfection

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Drying Consumables required: ▪ Cloths for wiping the medical device. ▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow the disinfectant to act according to the instructions of the disinfectant manufacturer.
  • Page 20: Servicing With Kavo Spray

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Care products and systems - Servicing Note KaVo guarantees the proper function of KaVo products only if the care products lis- ted as accessories are used, since these were tested for proper use on our products.
  • Page 21: Servicing With Kavo Sprayrotor

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Care products and systems - Servicing 7.5.2 Servicing with KaVo SPRAYrotor KaVo recommends servicing the product as part of the reprocessing after each use, i.e.
  • Page 22: Packaging

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 7 Reprocessing steps in accordance with ISO 17664 | 7.6 Packaging (no validated cleaning of the interior in accordance with German RKI requirements) KaVo recommends servicing the product as part of the reprocessing after each use, i.e.
  • Page 23: Sterilisation

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 7 Reprocessing steps in accordance with ISO 17664 | 7.7 Sterilisation ▶ Seal the medical device individually in the sterilised item packaging. 7.7 Sterilisation Sterilisation in a steam steriliser (autoclave) in accordance with EN...
  • Page 24: Tools And Consumables

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 8 Tools and consumables 8 Tools and consumables Available from dental suppliers. Material summary Mat.No. Instrument stand 2151 0.411.9501 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862...
  • Page 25: Terms And Conditions Of Warranty

    Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741 9 Terms and conditions of warranty 9 Terms and conditions of warranty The following warranty conditions apply to this KaVo medical device: KaVo provides the end customer with a warranty of proper function and guarantees...

This manual is also suitable for:

Smartmatic prophy s19 k1.011.67401.011.6741

Table of Contents