Hologic ThinPrep Imager Duo Operator's Manual page 215

Imaging station
Table of Contents

Advertisement

10
R
EMOTE
CAUTION
Network Configuration
Route all connectors
The Server monitor, keyboard, mouse, printer and a multiport Ethernet hub
carefully to avoid
must be located in the same area, so that interconnection cables easily reach
pinching the cables.
each component (within 1.8 meters [6 feet] of each other). Refer to Figure 10-1.
To avoid tripping
Review Scopes connected to the Server may be located up to 100 meters
over or
(328 feet) away, or extended another 100 meters using a second hub.
disconnecting
cabling, do not
place cabling near
foot traffic
10.2
Imager Duo Operator's Manual
S
S
U
'
ITE
ERVER
SER
S
Switch
Printer
P
Server*
Mouse
P
Keyboard
Monitor
Table, Desk or Lab Bench
Figure 10-1 Local Network Interconnection Schematic (Example)
G
UIDE
Up to 328 ft/
100 meters
P
P
Uninterruptable
Power Supply* and
Battery Backup
= Power Cord, Requires an Outlet
P
*May go on the floor, provided
there is no dust accumulation on or
around it.
P
Review
Scope
Review
Scope
Additional Scopes
depending on site
configuration
P