Hologic ThinPrep Imager Duo Operator's Manual page 194

Imaging station
Table of Contents

Advertisement

Table 6.2: Imaging Station and IPC Errors
Error #
4619
System Error. Consecutive black
frames detected (multiple slides).
4756
Slide has been rejected.
4757
4758
Slide has been rejected.
Insufficient fine focus operation.
4759
Three 4758 errors within a cas-
sette generates a 4759 error.
5050
Three consecutive 4450, 4451,
4452 errors (in any combination)
within a cassette generates a
5050 error.
5051
Three consecutive 4453, 4454,
4455 errors (in any combination)
within a cassette generates a
5051 error.
5073
Door(s) Open.
T
ROUBLESHOOTING
Error Text
Power cycle the system. If the error
persists, contact Technical Sup-
port.
• See Section C, Slide Preparation
• Check whether the cell spot is
• Check whether the slide is
See Section C, Slide Preparation
and Quality.
See errors 4450 through 4475 for
details.
• Follow the instructions on the
• Check that the doors are closed.
• At User Interface, select CON-
• If the problem is not resolved,
Imager Duo Operator's Manual
6
Corrective Action
and Quality.
thick or uneven.
uneven.
System monitor. (This may
require power cycling the Sys-
tem.)
TINUE in the User Assisted
Recovery dialog. If the problem
is resolved, continue processing.
power cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
6.13