Hologic ThinPrep Imager Duo Operator's Manual page 216

Imaging station
Table of Contents

Advertisement

CAUTION
Extended Network Configuration
Route all connectors
The Review Scopes may be located at a greater distance from the Server by
carefully to avoid
using an extended network interconnection between the Server hub/switch
pinching the cables.
and the Review Scope hub/switch.
To avoid tripping
over or
Low-level, industry standard interconnection devices such as bridges, repeat-
disconnecting
ers, media converters, copper and fiber wiring may be used. Refer to Figure 10-
cabling, do not
2. The distance that Review Scopes may be located away from the Server is
place cabling near
determined by the requirements of the specific apparatus used to create the net-
foot traffic.
work. The throughput of the extended network must be capable of 10Mbps or
greater.
Note:
P
P
Monitor
Table, Desk or Lab Bench
R
EMOTE
High-level interconnecting devices (routers and gateways) and wireless
devices must not be used. The devices used for the extended network
must form an isolated network used exclusively for the Imaging System
network. No other company networks, sub-networks or devices may be
connected to the network.
Network Interconnections
Switch
P
Printer
Server*
P
Mouse
Uninterruptable
Power Supply * and
Battery Backup
Keyboard
P
Figure 10-2 Extended Network Interconnection Schematic (Example)
S
S
U
ITE
ERVER
SER
P
Switch
P
= Power Cord, Requires an Outlet
* May go on the floor, provided there is no dust
accumulation on or around it.
Imager Duo Operator's Manual
10
'
G
S
UIDE
Up to 328 ft/
100 meters
P
Review
Scope
Review
Scope
Additional Scopes
depending on site
configuration
10.3