Hologic ThinPrep Imager Duo Operator's Manual page 129

Imaging station
Table of Contents

Advertisement

4
O
PERATION OF THE
Import Contents
Import date
Importing lab
Summary
Total slides
Slides imaged successfully
Slides with imaging errors
Minimum slide ID
Maximum slide ID
Activity and status:
Checking media
Preparing data
Updating database
Complete
To see a report of the import results, click the Results button. Refer to Figure 4-
46.
A new slide is new to the database.
A duplicate is a slide record that is already in the database and the slide data
being imported is identical. The slide data being imported is not written to the
database.
An error is a slide that has an error within its data - it will not have been
imported into the database.
A conflict is a condition where a slide record is already in the database and has
differing values from the slide record being imported. The slide data is not
imported and the Error ID is reported as a 'Conflict'. This might be a case of Lab
ID or Imager ID conflicts, but it is most often two different slides having the
same ID number.
The patient information must be checked and reconciled for both slides.
4.30