Hologic ThinPrep Imager Duo Operator's Manual page 195

Imaging station
Table of Contents

Advertisement

6
T
ROUBLESHOOTING
6.14
Imager Duo Operator's Manual
Table 6.2: Imaging Station and IPC Errors
Error #
5096
Fail connection to server.
6615
Slide has been rejected.
6616
Three consecutive 6615 errors
within a cassette generate a 6616
error.
6617
Slide has been rejected.
6618
Three consecutive 6617 errors
within a cassette generate a 6618
error.
6619
Vibration Error.
6628
Slide has been rejected.
6629
Three consecutive 6628 errors
within a cassette generate a 6629
error.
6630
Slide has been rejected.
6631
Three consecutive 6630 errors
within a cassette generate a 6631
error.
Error Text
• Check switch connections and
• Power Cycle the Imaging Sys-
This is a 'biological event' where
the sample may contain abundant
clumpy inflammation or bacteria.
• See Section C, Slide Preparation
• Manually review the slide.
See Section C, Slide Preparation
and Quality.
• The imaging station is sensitive
• Be aware of other things (slam-
This is a 'biological event' where
the sample may be too dense.
• See Section C, Slide Preparation
• Manually review the slide.
If the error persists, contact Tech-
nical Support.
Corrective Action
data bus harness connections
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
and Quality.
to vibrations. Look for vibrating
equipment (such as a vortexer)
near the imaging station.
ming doors, high traffic area)
that may also cause the vibra-
tions.
and Quality.