Hologic ThinPrep Imager Duo Operator's Manual page 87

Imaging station
Table of Contents

Advertisement

Figure 3-22 Interrupt Slide Processing Option
3. Use the mouse to select the option for when the batch is interrupted.
"Interrupt the current slide immediately (lose cassette progress)"
Use this option to stop batch processing at the current slide. The slide
will not be completed. The slide will be put away and then the Imaging
Station mechanisms will come to an orderly halt. The doors will unlock.
The batch is terminated and the slide data is not recorded.
"Interrupt the current slide immediately (save cassette progress)"
Use this option to stop batch processing at the current slide. The slide
will not be completed. The slide will be put away and the Imaging
Station mechanisms will come to an orderly halt. The doors will unlock.
If the batch is resumed (without a power cycle), the verification slide is run,
then the data for any previous slides in the cassette are written to the
Server and then the system continues processing at the slide that was in
progress during the interrupt.
"Interrupt at the conclusion of the current cassette"
Use this option to let the current cassette complete, run the verification
slide and write the slide data record information to the Server. The
slides are put away and the Imaging Station mechanisms will come to
an orderly halt. The doors will unlock.
If the batch is resumed, the system begins processing at the next
unprocessed cassette.