Hologic ThinPrep Imager Duo Operator's Manual page 196

Imaging station
Table of Contents

Advertisement

Table 6.2: Imaging Station and IPC Errors
Error #
6716
Failed to get a valid batch ID from
the server.
6901 through
Verification slide and focus errors.
6936
6962
Accelerometer fail.
6963
Fiducial Mark failure.
7001
Light Level Drift Outside of Toler-
ance.
T
ROUBLESHOOTING
Error Text
Power cycle the system. If the error
persists, contact Technical Sup-
port.
• See Section C, Slide Preparation
• Clean verification slide.
• If more than one Imaging Station
• Power Cycle the Imaging Sys-
• The imaging station is sensitive
• Be aware of other things (slam-
• Clean verification slide.
• Power Cycle the Imaging Sys-
• Clean verification slide.
• May be temperature related.
• Power Cycle the Imaging Sys-
Imager Duo Operator's Manual
6
Corrective Action
and Quality.
is in the lab, insure that the veri-
fication slides are not switched
among Imaging Stations.
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
to vibrations. Look for vibrating
equipment (such as a vortexer)
near the imaging station.
ming doors, high traffic area)
that may also cause the vibra-
tions.
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
Verify that the instrument is not
near a heating or A/C duct.
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.