Error #
2407
Slide has been rejected.
2409
Three consecutive 2407 errors
within a cassette generate a
2409.
3106
Slide not found.
3107
Three consecutive 3106 errors
within a cassette generate a
3107.
3111
Slot has multiple slides.
3113
Slot below has slide out of posi-
tion or multiple slides.
3159
Map Cassette Timeout.
3303
OCR Read Failed.
3304
3305
Invalid Slide ID.
3306
T
ROUBLESHOOTING
Error Text
• See Section C, Slide Preparation
• If more than one Imaging Station
• Ensure that the slide was prop-
• If slide was loaded correctly, see
• Clean cassette mapping sensor,
• Power Cycle the Imaging Sys-
• See Section C, Slide Preparation
• If label was printed on a T3,
• See Section C, Slide Preparation
• Call Tech Support to verify the
Imager Duo Operator's Manual
6
Corrective Action
and Quality.
is in the lab, insure that the veri-
fication slides are not switched
among Imaging Stations.
erly loaded in the cassette – with
label facing outward and only
one slide per slot.
Section C, Slide Preparation and
Quality.
refer to page 5.7.
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
and Quality.
replace ribbon.
and Quality.
Slide ID has the proper CRC.