Hologic ThinPrep Imager Duo Operator's Manual page 65

Imaging station
Table of Contents

Advertisement

2
I
MAGER
placed on the floor near the work area, provided it has adequate air circulation
to prevent dust accumulation and it is safely removed from foot traffic or other
interference. It should be accessible for routine maintenance.
The network printer will be installed and set up at the time of instrument instal-
lation. Please refer to the manufacturer's guide for operation and maintenance.
Note:
WARNING
The uninterruptable power supply will be installed and setup at the time of
instrument installation. Please refer to the manufacturer's guide for proper
Grounded Outlet
operation, maintenance and disposal.
SECTION
• The Imaging Station and the Image Processor Controller are sensitive to
• During operation, the Imaging Station is sensitive to vibrations. It should
2.4
Imager Duo Operator's Manual
D
S
I
UO
YSTEM
NSTALLATION
The uninterruptable power supply (UPS) provides battery backup
power for a period of time (up to 30 minutes for a fully charged battery)
in the event of a power outage. Slides will be stored and the robotic arm
is safely parked. The operator may then shut down the Imaging System
(refer to page 3.29).
STORAGE AND HANDLING - POST INSTALLATION
D
Environmental Considerations
sudden thermal or humidity changes. Do not locate them next to win-
dows, heaters, air conditioners, HVAC vents or doors that are frequently
opened and shut.
be placed on a sturdy flat surface away from centrifuges, vortexors or
any equipment that may cause vibrations. Keep away from other envi-
ronmental activity, such as constant foot traffic, proximity to elevators or
doors that are frequently open and shut.