Power Supply - Endress+Hauser HART Proline Prosonic Flow B 200 Operating Instructions Manual

Ultrasonic flow measuring system
Hide thumbs Also See for HART Proline Prosonic Flow B 200:
Table of Contents

Advertisement

Proline Prosonic Flow B 200 HART
Protocol-specific data
Terminal assignment
Supply voltage
Power consumption
Current consumption
Power supply failure
Electrical connection
Potential equalization
Terminals
Cable entries
Cable specification
Endress+Hauser
• For information on the device description files (® ä 45)
• For information on the dynamic variables and measured variables (HART device variables)
(® ä 45)
17.5

Power supply

Order characteristic for
"Output"
Option C
1) When using only one output: use output 1.
An external power supply is required for each output.
Order characteristic for
"Output"
Option C
1) External supply voltage of the power supply unit with load (® ä 21)
• Operation with output 1: 65 to 660 mW
• Operation with output 1 and 2: 130 to 1 320 mW
3.6 to 22 mA
If the option Defined value is selected in the Failure mode parameter (® ä 96):
3.59 to 22.5 mA
• Totalizers stop at the last value measured.
• Configuration is retained in the device memory (HistoROM).
• Error messages (incl. total operated hours) are stored.
(® ä 22)
No special measures for potential equalization are required.
Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm
• Cable gland: M20 × 1.5 with cable Æ 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry: NPT ½", G ½"
• Permitted temperature range: –40 °C (–40 °F) to ³ 80 °C (176 °F)
Minimum requirement: cable temperature range ³ ambient temperature + 20 K
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
Terminal numbers
Output 2
1)
3 (+)
4 (-)
4-20 mA
Minimum terminal voltage
1)
For 4 mA: ³ DC 18 V
For 20 mA: ³ DC 14 V
2
(20 to 14 AWG)
Technical data
Output 1
1 (+)
2 (-)
4-20 mA HART
Maximum terminal voltage
DC 30 V