Commissioning
58
100% bargraph value 3
Enter the 100% value to be
shown on the bargraph
display for the measured
value 3.
Value 4 display
Select the measured value
that is shown on the local
display.
10.4.5
Configuring the output conditioning
The Output conditioning wizard guides you systematically through all parameters that have to be
set for configuring the output conditioning.
Navigation path
"Setup" menu ® Output conditioning
Sequence of the wizard
å 15
"Output conditioning" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Display damping
Set the reaction time of the
local display to fluctuations in
the measured value caused
by process conditions.
Floating point number with
sign
• Volume flow
• Corrected volume flow
• Corrected methane
volume flow
• Energy flow
• Methane fraction
• Gross calorific value
• Wobbe index
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Current output 1
• Current output 2
Display damping
Damping out. 1
Damping out. 2
Mode output 1
Mode output 2
Forward flow
Forward/Reverse
End of wizard
Description
0 to 999 s
Proline Prosonic Flow B 200 HART
Depends on country and
nominal diameter
None
Rev. flow comp.
Selection/
Factory setting
User entry
0 s
Endress+Hauser
A0015993-EN
Need help?