Endress+Hauser HART Proline Prosonic Flow B 200 Operating Instructions Manual page 85

Ultrasonic flow measuring system
Hide thumbs Also See for HART Proline Prosonic Flow B 200:
Table of Contents

Advertisement

Proline Prosonic Flow B 200 HART
Endress+Hauser
261
Electronic modules
262
Module connection
270
Main electronic failure
271
Main electronic failure
272
Main electronic failure
273
Main electronic failure
275
I/O module failure
276
I/O module failure
282
Data storage
283
Memory content
311
Electronic failure
311
Electronic failure
*
Diagnostic behavior can be changed: Section 12.4 "Adapting diagnostic behavior"
Diagnostic
Short text
number
Diagnostics for the configuration
410
Data transfer
411
Up-/download active
431
Trim
437
Incompatible
configuration
438
Dataset
441
Current output 1-2
452
Calculation error
1. Restart device.
2. Check electronic modules.
3.Change I/O Modul or main
electronics.
1. Check module connections.
2. Change electronic modules.
Change main electronic module.
1. Restart device.
2. Change main electronic module.
1. Restart device.
2. Contact service.
1. Emergency operation via display.
2. Change main electronics.
Change I/O module.
1. Restart device.
2. Change I/O module.
1. Restart device.
2. Contact service.
1. Transfer data or reset device.
2. Contact service.
1. Transfer data or reset device.
2. Contact service.
Maintenance required!
1. Do not perform reset.
2. Contact service.
Remedial measures
1. Check connection.
2. Retry data transfer.
Up-/download active, please wait
Carry out trim.
1. Restart device.
2. Contact service.
1. Check data set file.
2. Check device configuration.
3. Up- and download new
configuration.
1. Check process.
2. Check current output settings.
1. Check device configuration.
2. Check process conditions.
Diagnostics and troubleshooting
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
M
Warning
Status
signal
Diagnostic behavior
from the
from the factory
factory
F
Alarm
C
Warning
C
Warning
F
Alarm
M
Warning
*
S
Warning
*
S
Alarm